Problem Description
Dr. Chen called, says the sterilizer is aborting cycles with pressure error 101. Needs urgent attention.
Technician Notes & Activity
D
Dispatch
Dec 19, 2025 at 12:23 PM
Work Order created from customer call.
A
Alice
Dec 20, 2025 at 12:23 PM
Initial diagnosis complete. Gasket is worn. Ordering part GKT101.
A
Alice
Dec 20, 2025 at 12:23 PM
Part GKT101 ordered.
A
Alice
Dec 20, 2025 at 12:23 PM
Issued loaner sterilizer LNR-M11-002 to customer.
A
Alice
Dec 20, 2025 at 12:23 PM
Status changed to On Hold pending parts.